Simultaneous determination of chromium(III) and chromium(VI) in aqueous solutions by ion chromatography and chemiluminescence detection

Research output: Contribution to journalJournal articleResearchpeer-review

A method for the simultaneous determination of chromium(iii) and chromium(vi) in a flow system based on chemiluminescence was developed. A Dionex cation-exchange guard column was used to separate chromium(iii) from chromium(vi), and chromium(vi) was reduced by potassium sulfite, whereupon both species were detected by use of the luminol-hydrogen peroxide chemiluminescence system. Linear calibration for both species was established over the concentration range 1-1000 micrograms l-1. The precision at the 20 micrograms l-1 level was 3.5% for chromium(iii) and 3.3% for chromium(vi), respectively. The detection limit was 0.5 micrograms l-1 for both species. Data were in agreement with Zeeman-effect background corrected atomic absorption spectrometry measurements.
Original languageEnglish
JournalAnalyst
Volume117
Issue number3
Pages (from-to)637-40
Number of pages4
ISSN0003-2654
Publication statusPublished - 1992

    Research areas

  • Chromatography, High Pressure Liquid, Chromium, Luminescence, Solutions, Trace Elements, Water